Description
Model CPB6000DP
Applications
- High-level primary standard
- Test and calibration of differential sensors and transmitters
- Cross-float measurement for piston-cylinder system effective area determination
- Positive or negative differential pressure
Special Features
- Measuring range (= static pressure + differential pressure) up to 800 bar / 11,600 psi pneumatic
- DP from 10 Pa up to 800 bar / 0.04 inH2O up to 11,600 psi with static pressure effects down to 1 ppm
- COFRAC calibration certificate included as standard
- High long-term stability with recommended recalibration cycle every five years
- Total measurement uncertainty to 0.008 % of differential pressure plus 0.0001 % (1 ppm) of static pressure